¡Atención! La norma vigente es:
INTE/ISO 80601-2-61:2023
INTE/ISO 80601-2-61:2020
Equipos electromédicos. Parte 2-61: Requisitos particulares para la seguridad básica y características de funcionamiento esenciales de pulsioxímetros para uso médico.
Objeto y campo de aplicación
Replacement:
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of PULSE OXIMETER EQUIPMENT intended for use on humans, hereafter referred to as ME EQUIPMENT. This includes any part necessary for NORMAL USE, including the PULSE OXIMETER MONITOR, PULSE OXIMETER PROBE, and PROBE CABLE EXTENDER.
These requirements also apply to PULSE OXIMETER EQUIPMENT, including PULSE OXIMETER MONITORS, PULSE OXIMETER PROBES and PROBE CABLE EXTENDERS, which have been REPROCESSED.
The intended use of PULSE OXIMETER EQUIPMENT includes, but is not limited to, the estimation of arterial oxygen haemoglobin saturation and pulse rate of PATIENTS in professional healthcare institutions as well as PATIENTS in the HOME HEALTHCARE ENVIRONMENT and the EMERGENCY MEDICAL SERVICES ENVIRONMENT.
This document is not applicable to PULSE OXIMETER EQUIPMENT intended for use in laboratory research applications nor to oximeters that require a blood sample from the PATIENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 201.11 and in 7.2.13 and 8.4.1 of the general standard.
NOTE 1 See also 4.2 of the general standard. “The general standard” is IEC 60601-1:2005+AMD1:2012, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
This document can also be applied to ME EQUIPMENT and their ACCESSORIES used for compensation or alleviation of disease, injury or disability.
This document is not applicable to PULSE OXIMETER EQUIPMENT intended solely for foetal use.
This document is not applicable to remote or slave (secondary) equipment that displays SpO2 values that are located outside of the PATIENT ENVIRONMENT.
NOTE 2 ME EQUIPMENT that provides selection between diagnostic and monitoring functions is expected to meet the requirements of the appropriate document when configured for that function.
This document is applicable to PULSE OXIMETER EQUIPMENT intended for use under extreme or uncontrolled environmental conditions outside the hospital environment or physician’s office, such as in ambulances and air transport. Additional standards can apply PULSE OXIMETER EQUIPMENT for those environments of use.
This document is a particular standard in the IEC 60601-1 and ISO/IEC 80601 series of standards.
Información general
Código del comité
EN
Nombre del comité
Endoso
Sector
Salud
ICS
11.040.40
Correspondencias
ISO 80601-2-61:2017
Organismos
ISO
Edición
1
Fecha de aprobación
2020-04-07
Número de páginas
118
Estado
Sustituido
Reemplazada por
Normas de Referencia
IEC 60601-1:2005+AMD1:2012, Clause 2 applies, except as follows: Replacement: IEC 60417, IEC 60529:2013, IEC 60601-1-2:2014, IEC 60601-1-6:2010 + Amendment 1:2013 IEC 60601-1-8:2006 + Amendment 1:2012 IEC 60825-1:2014, ISO 3744:2010, ISO 4135:2001, ISO 7000, ISO 7010, ISO 14155:2011, ISO 14937:2009, ISO 15223-1:2016, IEC 60068-2-27:2008, + Amendment 1:2013 IEC 60068-2-31:2008, IEC 60068-2-64:2008, IEC 60601-1-11:2015, IEC 60601-1-12:2014, IEC 60825-2:2004 + Amendment 1:2006+ Amendment 2:2012, IEC 62366-1:2015.